Online Phosphate Phosphorus Analyzer, HAPO-4000, is a measuring equipment designed to measure phosphate phosphorus by using absorption spectrophotometry (ascorbic acid reduction method) .HAPO-4000 performs continuous automatic measurement of phosphate phosphorus concentration in water based on the Standard Water Pollutant Testing Method, and it has wide range of applications in continuous automatic measurement and real-time monitoring device for management of drinking water quality, wastewater, industrial effluent, and groundwater.
Characteristics
HAPO-4000
Multi-channel valve system
-
- Patent acquired
Fixed quantity distribution of sample and reagent
- Patent acquired
Exactitude & Accuracy
Exactitude
-
-
- High performances micro pulse type pump
-
Accuracy
-
-
- Equipped with measurement wavelength selection option according to site conditions (880 or 710nm)
-
Immediacy & Stability
Immediacy
-
-
- Response within 15 minutes
-
Stability
-
-
- Built-in low-temperature reagent storage device for reagent stability
-
Convenience in operation
Convenience
-
-
- 3-point calibration and auto calibration function
- Easy and simple operation, self diagnosis
-
Operation
-
-
- Easy maintenance by simplification of the flow channel
- Easy operation by private-labeled operation program
-
Principle
When the sample flows in to the analyzer, the ammonium phosphorus molybdate that is created through reaction of phosphate ion in the sample with ammonium molybdate is reduced by ascorbic acid (C6H8O6)to generate the color of molybdenum blue. Here, phosphate phosphorus is quantified by measuring the blue absorbance of molybdenum blue at 880nm. Depending on the sample water conditions of the site, the measurement wavelength can be set to 710nm.
Specification
-
- Measurement item
Phosphate-Phosphorus (PO4-P)
-
- Measurement method
Absorbance spectrophotometry (Ascorbic acid reduction method)
-
- Measurement range
0~20mg/L (Adjustable)
-
- Wavelength
880nm (Option: 710nm)
-
- Measurement interval
15 minutes (Adjustable)
-
- Detection limit
0.01mg/L F.S.
-
- Accuracy
±2% F.S.
-
- Reproducibility
±2% F.S.
-
- Linearity
±2% F.S.
-
- Zero/Span drift
±2% F.S.
-
- Calibration
Auto/manual calibration by standard solution
-
- Sample/Reagent measuring method
Step motor pulse control
-
- Sample/Reagent distribution method
7-port distribution control method (Fixed quantity injection with multi-channel valve system)
-
- Operation system
Windows 7 (CPU process) or more
-
- Output
TCP/IP, DC 4~20mA, RS-232C
-
- Status display
Normal operation, on calibration, error
-
- Display
Color LED 10.4″ (Touch screen)
-
- Power
AC 220V, 50/60Hz
-
- Power consumption
Average 70W
-
- Voltage regulation
3% or less
-
- Data storage capacity
HDD 100G or more
-
- Data output
USB terminal
-
- Enclosure / Size
Steel, 1.5T(Except paint) / 645(W) x 439(D) x 1600(H) mm
-
- Operating humidity
RH 95% or less
Application
Application Field
Surface water monitoring
-
- River, lake, river water quality
- Waterworks facilities (intake station and Water plant, etc.)
Water quality TMS
-
- Sewage treatment plant, Wastewater treatment plant
- Wastewater discharging facility, Rural sewage treatment facility, etc.