Key Facts
- Frequency range from 20 Hz to 3.6/8/26.5/40 GHz
- 28 MHz signal analysis bandwidth, optionally 120 MHz
- Outstanding RF characteristics:
- TOI typ. +25 dBm
- 1 dB compression point +13 dBm
- 84 dBc ACLR/3GPP with noise correction
- –173 dBm (1 Hz) displayed average noise level (DANL) with noise correction and R&S®FSU-B24 preamplifier
- Analysis of signals in the analog and digital baseband
- Numerous firmware applications extending the signal analysis capabilities
- 16 Msample I/Q memory, expandable up to 705 Msample
Features & Benefits
-
Excellent RF performance for demanding applications
- Outstanding RF features
- 84 dB ACLR for 3GPP with noise correction
- 77 dB ACLR for 3GPP multicarrier signals (four adjacent carriers)
- TOI > +20 dBm, typ. +25 dBm
- 1 dB compression +13 dBm
- Displayed average noise level (DANL) typ. –173 dBm using noise correction (1 Hz bandwidth) and R&S®FSU-B24 preamplifier
- Phase noise –160 dBc (1 Hz) at 10 MHz carrier offset
- Phase noise –133 dBc (1 Hz) at 10 kHz carrier offset
- Outstanding RF features
-
Signal and spectrum analysis in a single instrument
- Signal analysis with up to 120 MHz bandwidth
- Signals from RF, analog or digital baseband
- Up to 705 Msample (I and Q)
- Low phase/frequency response
- User-definable sample rate
- Best signal fidelity for modulation analysis
- Ready for analysis of multistandard, multicarrier transmitters
- Signal analysis with up to 120 MHz bandwidth
-
Wide range of functions
With its wide range of functions, the R&S®FSQ is practically unparalleled on the spectrum analyzer market. Even the base unit includes all important functions.
- Channel filters from 100 Hz to 5 MHz
- RRC filters
- 1 Hz to 50 MHz resolution bandwidth (RBW)
- Number of measurement points/trace selectable between 155 and 30 001
- Time-selective spectrum analysis with gating function
- Up to 80 measurements/s in manual mode
- SCPI-compatible GPIB command set
- GPIB command set compatible with the R&S®FSE/R&S®FSIQ and legacy HP spectrum analyzer
- Fast ACP measurement in time domain
- Statistical signal analysis with CCDF function
- RMS detector
- Transducer factor for correcting antenna or cable frequency responses
- Harmonic distortion measurement function
- Spurious emission measurements with up to 100 001 points, in up to 20 segments
- External reference from 1 MHz to 20 MHz in 1 Hz steps
- LAN interface 100BaseT (1 Gbit)
- 16 Msample I and Q memory
Brief Description
The R&S®FSQ is the solution for all development and production measurement tasks. It offers very low phase noise, unsurpassed low residual EVM, a wide dynamic range and above-average accuracy, making it the ideal high-end measuring instrument for development applications, where tolerances and limit values must often be lower than defined in a standard.
Specifications
R&S®FSQ3 R&S®FSQ8 R&S®FSQ26 R&S®FSQ40 Frequency range 20 Hz to 3.6 GHz 20 Hz to 8 GHz 20 Hz to 26.5 GHz 20 Hz to 40 GHz Reference frequency aging: 1 x 10 –7/year, with R&S®FSU-B4 option: 2 x 10 –8 Spectral purity Phase noise typ. –133 dBc (1 Hz) at 10 kHz carrier offset, carrier frequency 500 MHz Distortion 1 Hz Sweep time Span >10 Hz 2.5 ms to 16 000 s Span 0 Hz (zero span) 1 μs to 16 000 s RBW 10 Hz to 50 MHz, FFT filter: 1 Hz to 30 kHz, channel filter, EMI bandwidth VBW 1 Hz to 30 MHz Display range DANL to +30 dBm DANL (1 Hz RBW) 1 GHz typ. –158 dBm typ. –155 dBm typ. –156 dBm typ. –155 dBm 7 GHz − typ. –154 dBm typ. –155 dBm typ. –153 dBm 13 GHz − − typ. –151 dBm typ. –151 dBm 26 GHz − − typ. –146 dBm typ. –147 dBm 40 GHz − − − typ. –141 dBm DANL with preamplifier ON
(R&S®FSU-B25), 1 GHz, 1 Hz RBW–162 dBm –162 dBm –162 dBm –162 dBm Displayed averagenoise level with preamplifier ON
(R&S®FSU-B24), 26 GHz, 10 Hz RBW− − typ. –166 typ. –166 dBm DANL with preamplifier ON
(R&S®FSU-B23), 26 GHz, 1 Hz RBW− − –150 dBm − Trace detectors Max Peak, Min Peak, Auto Peak, Sample, RMS, Average, Quasi Peak Total measurement uncertainty, f < 3.6 GHz 0.3 dB Display linearity 0.1 dB (0 dB to –70 dB)